Future Science Group
Browse

Comparison of Bruker Biotyper (R) and Vitek (R) MS matrix-assisted laserdesorption/ionization–time-of-flight massspectrometry platforms for the identificationof filamentous fungi - Supplemental Table 1

Download (38.68 kB)
dataset
posted on 2023-06-15, 14:51 authored by Julia D Hankins, Megan H Amerson-Brown, Cameron A Brown, Ashleigh N Riegler, Kenneth L Muldrew, James J Dunn

Aims: To evaluate the performance of twomatrix-assisted laser desorption/ionization–time-of-flight mass spectrometry platforms to identify molds isolated from clinical specimens. Methods: Fifty mold isolates were analyzed on Bruker Biotyper R ? and Vitek R ? MS platforms. Two Bruker Biotyper extraction protocols were assessed alongside the US FDA-approved extraction protocol for Vitek MS. Results: The Bruker Biotyper modified NIH-developed extraction protocol correctly identified more isolates than Bruker’s protocol (56 vs 33%). For species in the manufacturers’ databases, Vitek MS correctly identified 85% of isolates, with 8% misidentifications. The Bruker Biotyper identified 64%, with no misidentifications. For

isolates not in the databases, the Bruker Biotyper did not misidentify any, and VitekMSmisidentified 36%. Conclusion: Both the VitekMS and Bruker Biotyper accurately identified the fungal isolates, however Vitek MS was more likely to misidentify isolates than the Bruker Biotyper.

History

Usage metrics

    Future Microbiology

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC